IXC458I
SIGNAL direction DEVICE devn STOPPED: text

Explanation

In the message, text is one of the following:
  • REASON UNKNOWN
  • OPERATOR COMMAND
  • RETRY LIMIT EXCEEDED
  • SYSPLEX PARTITIONING OF LOCAL SYSTEM
  • SYSPLEX PARTITIONING OF REMOTE SYSTEM
  • OTHER SIDE IS SAME DIRECTION
  • SUBCHANNEL NOT OPERATIONAL FOR RESUME
  • START REQUEST FAILED
  • CONNECTED TO NON-XCF SIGNALLER
  • NON-XCF SIGNALLER USING PATH
  • HALT I/O FAILED
  • PURGE I/O FAILED

Cross-system coupling facility (XCF) stopped a signalling path. Unless the operator initiated the stop, the signalling path remains defined to XCF.

In the message text:
direction
The direction of the signalling path that stopped. direction is one of the following:
  • PATHIN for an inbound signalling path.
  • PATHOUT for an outbound signalling path.
devn
The device number for the signalling path that stopped.
REASON UNKNOWN
XCF cannot determine why the signalling path was stopped.
OPERATOR COMMAND
The operator entered a SETXCF STOP command. The device is no longer defined to XCF.
RETRY LIMIT EXCEEDED
The retry count for the signalling path has exceeded the retry limit. The device might still be usable. The problem can be:
  • An I/O error occurred on the device.
  • An incorrect COUPLExx parmlib member was specified.
  • The specified COUPLExx parmlib member had signalling path definition errors. For instance, the retry limit can be exceeded if both sides of a signalling path were started in the same direction. See the explanation for the message text OTHER SIDE IS SAME DIRECTION.
  • The IPL of a system into a sysplex caused an ESCON® CTC device to appear non-operational.
SYSPLEX PARTITIONING OF LOCAL SYSTEM
The local system is no longer in the sysplex, so XCF stopped the signalling path connected to it. If the operator used either the VARY XCF command with RETAIN=NO or the SETXCF STOP command, the definition of the signalling path is deleted from XCF.
SYSPLEX PARTITIONING OF REMOTE SYSTEM
The remote system is no longer in the sysplex, so XCF stopped the signalling path connected to it. If the operator used either the VARY XCF command with RETAIN=NO or the SETXCF STOP command, the definition of the signalling path is deleted from XCF.
OTHER SIDE IS SAME DIRECTION
XCF tried to establish signalling connectivity between two systems, but the signalling path was defined in the same direction on both systems. A signalling path must have an outbound side and an inbound side. If both sides are defined in the same directions, messages cannot travel between the two systems involved.

Either of the two systems involved can detect the problem; the message appears only on one system. No response or acknowledgment is provided about this condition to the other system involved.

SUBCHANNEL NOT OPERATIONAL FOR RESUME
The subchannel is not operational for one of the following reasons:
  • No subchannel is provided.
  • The subchannel did not have a valid device number assigned.
  • The subchannel is not enabled.
START REQUEST FAILED
A request to start a device for use as a signalling path failed because the device is either not suitable or not available. See message IXC456I for an explanation of the problem.
CONNECTED TO NON-XCF SIGNALLER
The device on the other end of this signalling path is not defined to XCF. Either this signalling path is not connected to a system in the sysplex, or another application is trying to use the signalling path. This signalling path can only be used for communication between two systems in the sysplex.

This message can also be issued in a circumstance where XCF is using the link. If an ESCON CTC is defined to the hardware as a BCTC in the IOCDS, but defined to software as an SCTC, XCF will attempt to use the CTC. The protocol understood by the hardware for the CTC is for BCTC; therefore, the first system that tries to connect shows status "LINKING". However, when the second system attempts to connect, the system issues message IXC458I with the additional text CONNECTED TO NON-XCF SIGNALLER.

NON-XCF SIGNALLER USING PATH
An non-XCF application tried to use this system's signalling path. Devices used by XCF must be dedicated exclusively to XCF.
HALT I/O FAILED
The system tried to stop all I/O through this device, but the request failed. The device is probably in a permanent error state. If the device is an ESCON CTC, IPLing another system into the sysplex might have caused the device to stop.
PURGE I/O FAILED
An attempt to remove all I/O queued to this device failed.

System action

XCF stops using this device as a signalling path. The device is left unallocated and online. XCF writes a component trace record for the problem.

If the stop was not initiated by the operator, the device remains defined to XCF, and is in an inoperative state.

Note: Depending on the kind of failure, XCF may be able to automatically restart the device.

Operator response

Depending on the message text, do one of the following:
SUBCHANNEL NOT OPERATIONAL FOR RESUME
HALT I/O FAILED
Contact hardware support.
PURGE I/O FAILED
Contact software support.
OTHER SIDE SAME DIRECTION
The direction of the signalling path direction is wrong. Correct the error as follows:
  • Enter a SETXCF STOP direction to delete the definition of the incorrect path.
  • Enter SETXCF START direction with the correct direction.
  • For other message text, notify the system programmer.
  • Manually restart the path.

System programmer response

For both systems involved, do the following:
  • Make sure that the correct device number was specified.
  • Make sure that the signalling path definitions in the COUPLExx parmlib member are correct.
  • Make sure that the correct COUPLExx parmlib member was specified.
  • Make sure that the device is being used only by XCF.
  • Look in the SYS1.LOGREC error records for I/O errors on the device.
  • Ask the operator to enter DISPLAY XCF,PATHOUT,DEVICE=ALL or the DISPLAY XCF,PATHIN,DEVICE=ALL commands on the active systems for information on devices.

    If the problem persists, search problem reporting data bases for a fix for the problem. If no fix exists, contact the IBM® Support Center. Provide the XCF component trace data, the SYS1.LOGREC error record, and the GTF trace data for the device that has failed.

Source

Cross-system coupling facility (XCF)

Module

IXCT1STR

Routing code

1, 2, 10

Descriptor code

12