IV39969: DATA SHEET STATUS PASSES ON WORK ORDER WHEN CALIBRATION POINTS FAIL

Subscribe

You can track all active APARs for this component.

APAR status

  • Closed as user error.

Error description

  • Log in to maxdemo instance as wilson
    Go To Planning | Data Sheet Template
    Click on the New Data Sheet button on the toolbar
    In the Data Sheet field enter "DS-1"
    In the Asset Functions section click the New Row button
    
    In the Asset Function field enter 1; in the Description field
    enter
    "Asset Function 1"
    In the Input Information section enter values:
    Input Range From: 100.00
    To: 300.00
    Units: psi
    In the Output Information section enter values:
    Output Range From: 4.00
    To: 20.00
    Units: mA
    Click the checkbox for Tolerance on Output?
    In the Tolerance 1 section enter values:
    Description: Asset Function 1 / Tolerance 1
    Single From: -1.00
    To: 1.00
    As: EU
    Status when Exceeded: FAIL
    
    On the Calibration Points tab click the New Row button
    Enter values:
    Calibration Point: 1 Description: Asset Function 1 / Calibration
    Point 1
    Nominal Input: 100.00
    Click the New Row button and Enter Values:
    Calibration Point: 2 Description: Asset Function 1 / Calibration
    Point 2
    Nominal Input: 200.00
    
    Scroll back up and collapse the details for Asset Function 1 and
    click
    the Duplicate Row button
    Enter values:
    Asset Function: 2 Description: Asset Function 2
    To: 300.00
    (Tolerance 1) Description: Asset Function 2 / Tolerance 1
    On the Calibration Points tab click the New Row button
    Enter values:
    Calibration Point: 1 Description: Asset Function 2 / Calibration
    Point 1
    Nominal Input: 100.00
    Click the New Row button and Enter Values:
    Calibration Point: 2 Description: Asset Function 2 / Calibration
    Point 2
    Nominal Input: 200.00
    
    Collapse the details for the Calibration Point, scroll back up,
    and
    collapse the details for the Asset Function
    In the Asset Functions section, with Asset Function 2
    highlighted, click
    the Duplicate Row button
    Enter values:
    Asset Function: 3 Description: Asset Function 3
    (Tolerance 1) Description: Asset Function 3 / Tolerance 1
    On the Calibration Points tab click the New Row button
    Enter values:
    Calibration Point: 1 Description: Asset Function 3 / Calibration
    Point 1
    Nominal Input: 100.00
    Click the New Row button and Enter Values:
    Calibration Point: 2 Description: Asset Function 3 / Calibration
    Point 2
    Nominal Input: 200.00
    Save the record
    
    Click the Change Status button on the toolbar
    In the New Status field click Approved
    Click OK
    
    Go To Work Orders | Work Order Tracking(Cal)
    Click on the New Work Order button on the toolbar
    In the Asset field enter 13130 (or other valid asset record)
    In the Work Type field | Select Value | click CAL
    On the Data Sheet tab | Associated Data Sheets section | click
    the New
    Row button
    In the Data Sheet field enter "ds-1" and tab out
    Save the record
    Click the Change Status button on the toolbar
    In the New Status field click Approved
    Click OK
    
    On the Calibration Points tab enter values for Asset Function 1:
    Calibration Point 1  As Found Input 100.00  / As Found Output
    4.00    As
    Left Input 100.00  / As Left Output 4.00
    Calibration Point 2  As Found Input 100.00  / As Found Output
    4.00    As
    Left Input 100.00  / As Left Output 4.00
    Save the Work Order; Note that As Found Status and As Left
    Status
    updates to PASS
    
    Highlight Asset Function 2
    On the Calibration Points tab enter values for Asset Function 2:
    Calibration Point 1  As Found Input 102.00  / As Found Output
    4.50    As
    Left Input 102.00  / As Left Output 4.50
    Calibration Point 2  As Found Input 102.00  / As Found Output
    4.50    As
    Left Input 102.00  / As Left Output 4.50
    Save the Work Order; Note that As Found Status and As Left
    Status
    updates to PASS; should be FAIL
    
    Highlight Asset Function 3
    On the Calibration Points tab enter values for Asset Function 2:
    Calibration Point 1  As Found Input 102.00  / As Found Output
    4.50    As
    Left Input 102.00  / As Left Output 4.50
    Calibration Point 2  As Found Input 102.00  / As Found Output
    4.50    As
    Left Input 102.00  / As Left Output 4.50
    Save the Work Order; Note that As Found Status and As Left
    Status
    updates to PASS; should be FAIL?
    
    -In the Associated Data Sheets section | As Found Status and As
    Left
    Status display PASS; 2 of 3 Calibration Points have
    failed
    

Local fix

Problem summary

Problem conclusion

Temporary fix

Comments

  • Based on the steps, the calibration should pass.
    Output Range From: 4.00 To: 20.00
    The tolerance is -1 to 1 and in the steps, it stated that when
    setting as-left to 4.50, it should set the the status to fail
    4.50 is between 3.00 and 5.00 so it is a valid for it to PASS.
    If I set the value to 2.50, it would set the status to FAIL,
    which is correct.
    

APAR Information

  • APAR number

    IV39969

  • Reported component name

    MAXIMO CALIBRAT

  • Reported component ID

    5724R5700

  • Reported release

    712

  • Status

    CLOSED USE

  • PE

    NoPE

  • HIPER

    NoHIPER

  • Special Attention

    NoSpecatt

  • Submitted date

    2013-04-16

  • Closed date

    2013-07-01

  • Last modified date

    2013-07-01

  • APAR is sysrouted FROM one or more of the following:

  • APAR is sysrouted TO one or more of the following:

Fix information

Applicable component levels



Rate this page:

(0 users)Average rating

Document information


More support for:

IBM Maximo Calibration
Calibration

Software version:

712

Reference #:

IV39969

Modified date:

2013-07-01

Translate my page

Machine Translation

Content navigation